A high
HomeHome > Blog > A high

A high

May 25, 2023

Scientific Reports volume 12, Article number: 13798 (2022) Cite this article

1475 Accesses

1 Citations

2 Altmetric

Metrics details

Leukapheresis, the extracorporeal separation of white blood cells (WBCs) from red blood cells (RBCs) and platelets (PLTs), is a life-saving procedure used for treating patients with cancer and other conditions, and as the initial step in the manufacturing of cellular and gene-based therapies. Well-tolerated by adults, leukapheresis poses a significant risk to neonates and low-weight infants because the extracorporeal volume (ECV) of standard centrifugation-based machines represents a particularly large fraction of these patients’ total blood volume. Here we describe a novel high-throughput microfluidic device (with a void volume of 0.4 mL) based on controlled incremental filtration (CIF) technology that could replace centrifugation for performing leukapheresis. The CIF device was tested extensively using whole blood from healthy volunteers at multiple hematocrits (5–30%) and flow rates (10–30 mL/min). In the flow-through regime, the CIF device separated WBCs with > 85% efficiency and 10–15% loss of RBCs and PLTs while processing whole blood diluted with saline to 10% hematocrit at a flow rate of 10 mL/min. In the recirculation regime, the CIF device demonstrated a similar level of separation performance, virtually depleting WBCs in the recirculating blood (~ 98% reduction) by the end of a 3.5-hour simulated leukapheresis procedure. Importantly, the device operated without clogging or decline in separation performance, with minimal activation of WBCs and PLTs and no measurable damage to RBCs. Compared to the typical parameters of centrifugation-based leukapheresis, the CIF device had a void volume at least 100-fold smaller, removed WBCs about twice as fast, and lost ~ 2–3-fold fewer PLTs, while operating at a flow rate compatible with the current practice. The hematocrit and flow rate at which the CIF device operated were significantly higher than previously published for other microfluidic cell separation methods. Finally, this study is the first to demonstrate a highly efficient separation of cells from recirculating blood using a microfluidic device. Overall, these findings suggest the feasibility of using high-throughput microfluidic cell separation technology to ultimately enable centrifugation-free, low-ECV leukapheresis. Such a capability would be particularly useful in young children, a vulnerable group of patients who are currently underserved.

Leukapheresis is a complex medical procedure during which a patient's blood is passed through an apheresis machine to collect white blood cells (WBCs) and return red blood cells (RBCs) and platelets (PLTs) back to the patient1,2. Leukapheresis enables two potentially life-saving applications: leukodepletion and WBC collection. Leukodepletion can be used to reduce a dangerously high WBC count in patients with leukemia3,4, or to remove activated WBCs as a drug-free treatment for inflammatory bowel disease5,6,7 and other conditions8,9. Collection of WBCs via leukapheresis is the initial step in manufacturing a wide range of cellular therapies2, including granulocyte infusion10,11, adoptive immunotherapies12,13,14,15, hematopoietic stem cell transplantation16,17,18, and novel gene-based treatments19,20.

Although generally well-tolerated by adults and older children, performing leukapheresis for neonates and low-weight infants is technically challenging and clinically risky. Currently, leukapheresis is performed using centrifugation-based apheresis machines, which have a substantial extracorporeal volume (ECV) typically ranging 150–250 mL (excluding any additional tubing to connect to the patient)2,21, whereas total blood volume (TBV) of a 10 kg infant is only ~ 750 mL22,23. Because the ECV represents such a large fraction of their TBV, pediatric patients experience a significantly higher incidence of serious complications associated with the leukapheresis procedure, including hypotension, symptomatic hypocalcemia, allergic reactions, catheter-related thrombosis, infections, severe anemia, and even death21,24,25,26,27,28. In principle, WBCs can also be separated using a regular leukoreduction filter29,30, or a column packed with cellulose acetate beads that selectively adsorb activated granulocytes and monocytes6. However, the separation capacity of these devices is limited by their physical size, WBCs trapped by the devices continue to release proinflammatory cytokines into the bloodstream, and the recovery of trapped WBCs via backflush is rather modest6,29,30.

The earliest attempt to miniaturize leukapheresis using microfluidics utilized a continuous-flow diffusive filter to remove WBCs from a sample of blood with high efficiency (~ 97%), albeit with a significant RBC loss (~ 50%) and a flow rate of only 5 µL/min31. Since then, the use of microfluidic technologies for blood cell separation (although not specifically for leukapheresis) has steadily increased32. Devices based on ‘inertial focusing’33,34 can perform cell separation at relatively high flow rates (1–20 mL/min) but require significant dilution of the blood sample (< 1% hematocrit, HCT)35,36,37. Devices based on ‘deterministic lateral displacement’ (DLD)38 require significantly less dilution, but have exceedingly small minimal feature sizes (< 5 µm), require multiple pumps and extremely high driving pressures to operate at reasonable flow rates (< 7 mL/min), and often subject cells to shear stress sufficiently high to activate PLT and von Willebrand factor39,40,41,42. ‘Controlled incremental filtration’ (CIF) overcame these limitations by having a substantially larger minimal feature size (~ 20 µm), and thus reducing fluidic resistance and shear throughout the device43,44. As a result, a CIF-based device was able to separate > 85% of WBCs (with < 30% loss of RBCs and PLTs) from the mononuclear cell (MNC) concentrates (~ 5% HCT) at flow rates of up to 30 mL/min46 and PLTs concentrated and leukoreduced using CIF were minimally activated43,44. Notwithstanding these significant advancements, none of the aforementioned devices have been tested in the recirculation regime for any significant amount of time, and the effects of such processing on blood cells and device performance remain largely unknown.

In this study, we describe a new CIF-based device with a void volume of only 0.4 mL designed for high-throughput separation of WBCs from minimally diluted, recirculating whole blood. We first tested the separation performance of the device in a flow-through regime while adjusting multiple variables independently: HCT (5–30%), filtrate:retentate flow ratio (2–13:1), and flow rate (10–30 mL/min). The device demonstrated > 85% efficiency of WBC removal (with only 10–15% loss of RBCs and PLTs) while processing diluted whole blood with 10% HCT at a flow rate of 10 mL/min. When tested in the recirculation regime, the device operated without clogging or any decline in separation performance, with minimal activation of WBCs and PLTs and no measurable damage to RBCs over the entire duration of a 3.5-hour simulated leukapheresis procedure. The WBC concentration in the recirculating blood declined exponentially, decreasing ~ 98% by the end of the procedure.

The operating principle of controlled incremental filtration (CIF) and the computational framework used for generating CIF-based device designs have been described in detail previously43,44. A typical CIF design consists of three collinear flow channels separated by a series of ‘pill’-shaped posts, which define the filtration gaps connecting the central and the side channels fluidically (Fig. 1). As the blood sample flows through the central channel, a small fraction of that flow is siphoned off into the side channels through each filtration gap. The width of the side channels gradually increases to accommodate the influx of filtrate. The rate of this increase is calculated, as previously described43, to precisely control the fraction of flow extracted through each gap. It is the magnitude of this flow fraction—not the width of the gaps—that determines the size cutoff for the cells that are too large to be pulled into the side channels by the filtrate (‘critical diameter’). The critical diameter of the CIF design used in this study was 6 µm to retain most WBCs in the central channel while allowing RBCs and PLTs to outflow into the side channels unimpeded (Fig. 1). Because the flow fraction extracted through each gap is relatively small, any CIF design must incorporate thousands of gaps to achieve the desired filtrate:retentate flow ratio43,44. In practice, the number of filtration gaps (and, hence, the flow ratio) is limited by the maximum channel length allowed by the fabrication technique44,46.

The separation of WBCs from RBCs and PLTs via controlled incremental filtration (CIF). Most WBCs are too large to be pulled by the filtrate into the side channels, and therefore they are retained and concentrated in the central channel. RBCs and PLTs are smaller than the critical diameter of the device, and therefore they are distributed between the filtrate and the retentate according to the device flow ratio.

The CIF device comprised an array of 48 individual CIF elements multiplexed in parallel (Fig. 2). Each CIF element had an overall footprint of 1.4 mm × 73 mm and consisted of (Fig. 2a): (i) a built-in filter for retaining microaggregates and other debris that may be present in the blood sample, (ii) a transition region where each side channel comprised a series of progressively shorter serpentine segments to ensure the appropriate flow fraction was extracted while maintaining minimal feature size at ~ 20 µm44, and (iii) a linear separation region (~ 61 mm-long) in which the width of the side channels progressively increased and the width of the central channel decreased throughout its length. A series of through holes in the device layer provided fluidic access to the common inlet and the outlets of all CIF elements. The outputs of the central channels were collected through a network of large channels in the separate top layer (Fig. 2b). A fully assembled CIF device had one inlet through which the blood sample was distributed to each CIF element of the device, and two outlets for collecting the outputs of the central channels (retentate) and side channels (filtrate) of all CIF elements (Fig. 2c). The void volume of the multiplexed CIF device was 0.4 mL, excluding the tubing (Fig. 2c).

CIF-based microfluidic device. (a) Design of the CIF element. Width of the side channel near inlet ws(in) = 21 µm and near outlet ws(out) = 154 µm. Width of the central channel near inlet wc(in) = 120 µm and near outlet wc(out) = 70 µm. Filtration gap width g = 19 µm. Depth of all channels is 140 µm throughout. (b) Components of the multiplexed CIF device: (i) a ‘top layer’ containing a system of larger channels for collecting the central channel output (retentate) from each CIF element of the device, (ii) a ‘device layer’ consisting of individual CIF elements arranged in parallel, and (iii) a flat substrate for sealing the channels of the device. (c) An assembled microfluidic device comprised of 48 multiplexed CIF elements filled with a 10% HCT blood sample. Scale bar is 1 cm.

We first tested the effect of sample HCT on the CIF device performance. The efficiency of WBC removal was the highest (88.4 ± 1.3%) for the sample with 10% HCT and decreased as the HCT of the sample increased (Fig. 3a). For samples with 20% HCT, the CIF device was able to remove 57 ± 7% of the WBCs initially present in the blood passing through the device. The percent removal (loss) of RBCs and PLTs increased modestly with increasing HCT, from 10.6 ± 0.9% for RBCs and 9.1 ± 0.5% for PLTs at 5% HCT to 14.8 ± 0.1% for RBCs and 14.1 ± 0.5% for PLTs at 30% HCT (Fig. 3a). As expected for cells that are smaller than the critical diameter of the device, RBC and PLT percent losses closely matched the values predicted based on the device flow ratio, i.e. 100/(1 + flow ratio) (Fig. 3a, dashed line).

Performance of the CIF device in the flow-through regime. Dependence of cell removal on (a) sample HCT (10 mL/min flow rate, n = 3), (b) device flow ratio (10% HCT, 10 mL/min flow rate, n = 5), and (c) flow rate (10% HCT, n = 3). In panels (a) and (c), the flow ratio was set at 8.70 ± 0.39 and 8.21 ± 0.13 respectively, and then allowed to change depending on the parameters of the experiment. In each panel, a dashed line indicates the expected percent removal for cells that are smaller than the critical diameter of the device calculated based on the flow ratio measured in each experiment. Values shown are mean ± standard deviation.

A CIF device is designed to have a specific flow ratio, assuming the same pressure in the device outlets. To control the flow ratio in real-time, we created a difference in pressure between the outlets by varying the relative heights of the reservoirs collecting the retentate and filtrate. The actual flow ratio at which the device operated was then calculated by simply dividing the volume of the filtrate by the volume of the retentate collected from the device per unit of time. This simple manipulation allowed us to adjust the flow ratio within a wide range (~ sixfold, Fig. 3b). Lower flow ratios were associated with higher WBC removal, which remained consistently above 80% for all flow ratios up to ~ 9 (reaching 84 ± 3% at the flow ratio of 8.9 ± 1.5). The removal (loss) of RBCs and PLTs followed the expected reciprocal dependence on flow ratio (Fig. 3b, dashed line), decreasing rapidly from 37.5 ± 0.8% for RBCs and 36.1 ± 2.4% for PLTs at the flow ratio of 1.9 ± 0.1 (34% expected loss) to 10.9 ± 1.4% for RBCs and 11.4 ± 1.1% for PLTs at the flow ratio of 8.9 ± 1.5 (10% expected loss), and further down to 8.2 ± 0.9% for RBCs and 7.9 ± 1.4% for PLTs at the flow ratio of 12.6 ± 2.1 (7% expected loss) (Fig. 3b). Given such a strong dependence of RBC and PLT loss on device flow ratio, the ability to adjust the flow ratio in real-time could be useful for matching key parameters of the leukapheresis procedure (i.e., WBC removal, RBC and PLT loss) to the unique needs of individual patients.

We further evaluated the separation efficiency of the CIF device at flow rates ranging from 10 to 30 mL/min (Fig. 3c). WBC removal declined gradually with increasing flow rate from 88.0 ± 2.5% at 10 mL/min to 81.0 ± 1.4% at 30 mL/min. The flow ratio of the device also declined from 8.21 ± 0.13 at 10 mL/min down to 4.00 ± 0.03 at 30 mL/min (Fig. 3c). This decline was caused by the distortion of the CIF channel geometry (bulging) due to the elastic deformation of PDMS at higher driving pressures / flow rates, as we have previously observed for similar devices46. Consequently, the RBC and PLT loss increased from 12.0 ± 1.1% for RBCs and 13.1 ± 1.5% for PLTs at 10 mL/min to 19.4 ± 0.6% for RBCs and 18.0 ± 2.0% for PLTs at 30 mL/min, following the trend expected based on the flow ratio (Fig. 3c, dashed line).

Next, we tested the CIF device in the recirculation regime using the operational parameters (flow rate, flow ratio, sample HCT) that maximized device performance in the flow-through experiments. In our recirculation setup (Fig. 4a), a blood bag filled with 179.7 ± 0.8 mL of diluted WB (10% HCT) was used to emulate the TBV of a subject. During each recirculation round, 56.5 ± 0.8 mL of the blood sample was withdrawn from the bag (via a length of tubing inserted all the way to the bottom of the bag through one of its sampling ports) and then passed through the CIF device at a flow rate of 10 mL/min. The retentate (5.6 ± 0.5 mL per round) was collected into a conical tube, which was elevated above the CIF device to produce the desired flow ratio of 8.8 ± 0.3 (Fig. 4a). To calculate the flow ratio for a device operating in the recirculation regime, the volume of the filtrate was estimated as the difference between the volume pushed through the device and the volume of the retentate collected into the waste reservoir per recirculation round. The filtrate was returned to the bag through the other sampling port at the top of the bag, to effectively place the blood ‘inlet’ and ‘outlet’ at the opposite sides of the bag and thus promote mixing. Additionally, the blood bag was mixed during and after each recirculation round to ensure a uniform distribution of cells within the bag. This ‘withdraw-infuse’ cycle was repeated 12 times over ~ 3.5 hours. Following the completion of each round, a 0.2 mL sample was taken from the bag (via the sampling valve) and the retentate for measurements. The volumes of the retentate and each sample withdrawn were recorded to track the change in volume of the blood bag after each recirculation round.

Removal of WBCs from blood during recirculation. (a) Experimental setup used for the recirculation experiments. The blood bag was filled with 179.7 ± 0.8 mL of WB diluted to 10% HCT with normal saline. The waste reservoir collecting the retentate (concentrated WBCs) was placed 55 cm above the device; the resulting device flow ratio was 8.8 ± 0.3. At each recirculation round, 56.5 ± 0.8 mL of the sample was withdrawn from the bag and then infused through the CIF device at a flow rate of 10 mL/min, while producing 5.6 ± 0.5 mL of retentate (waste). A 1-mL syringe was used to sample the blood coming from the bag. Arrows indicate the direction of flow of blood in the circuit. (b) Changes in cell concentrations following each round of CIF-based cell separation in the recirculation regime. Values shown are mean ± standard deviation (n = 9, using blood from 6 unique subjects). Solid lines are linear fits for volume (y = − 0.0608x + 1.7975, R2 = 0.9997), RBCs (y = − 0.0044x + 1.1037, R2 = 0.8218) and PLTs (y = − 0.0059x + 0.6072, R2 = 0.8485), and a model fit for WBCs (with WBC removal of 81% minimizing the root-mean-square error).

The concentration of WBCs in the blood bag declined exponentially throughout the recirculation experiment (Fig. 4b). After 3 recirculation rounds (~ 170 mL processed volume), WBC concentration decreased by 55%, from 1.59 ± 0.41 × 103/µL in the initial sample down to 0.72 ± 0.26 × 103/µL. At round 6 (~ 340 mL processed volume), WBC concentration decreased further down to 0.28 ± 0.15 × 103/µL (82% decrease from the initial level). By the end of the recirculation experiment (round 12, ~ 680 mL processed volume), WBCs in the bag were virtually depleted (~ 98% decrease from the initial level). The concentration of RBCs and PLTs decreased linearly over the entire duration of the recirculation experiment (Fig. 4b): by ~ 4.5% for RBCs (from 1.09 ± 0.03 × 106/µL to 1.04 ± 0.05 × 106/µL; y = − 0.0044x + 1.1037, R2 = 0.8218), and by ~ 14.5% for PLTs (from 0.62 ± 0.07 × 105/µL to 0.53 ± 0.06 × 105/µL; y = − 0.0059x + 0.6072, R2 = 0.8485). Cells that are smaller than the critical diameter of the CIF device distribute according to the flow ratio, and therefore their concentration should be the same in both filtrate and retentate. Hence, we investigated the effect of blood recirculation through the CIF device on the properties of blood cells.

We used imaging flow cytometry (FC) to measure activation of WBCs (CD11b) and PLTs (CD62P), and PLT-WBC aggregate formation for samples collected from the bag before (round 0), during (round 6), and immediately after (round 12) the recirculation experiments (Fig. 5). Activation of separated WBCs (retentate output) was measured during (round 6) and after (round 12) the experiment. Additionally, in four out of five recirculation experiments, a sample of the subjects’ blood was set aside on the benchtop (BT) for the duration of the experiment to control for activation of WBCs and PLTs from simply being stored at room temperature for a prolonged period. All samples were tested at rest (as collected) and after incubation with either phorbol 12-myristate 13-acetate (PMA; for WBCs, Fig. 5a, b) or thrombin receptor agonist peptide-6 (TRAP; for PLTs and PLT-WBC aggregates, Fig. 5c, d) to evaluate how CIF recirculation affected the ability of WBCs and PLTs to become activated in response to relevant stimuli.

Effect of CIF processing in the recirculation regime on the activation state of WBCs and PLTs. (a) Activation of WBCs (CD11b) remaining in the recirculating blood. (b) Activation of WBCs (CD11b) removed with the retentate (waste). (c) Activation of recirculating PLTs (CD62P). (c) Formation of PLT-WBC aggregates in the recirculating blood. Measurements were performed via flow cytometry for samples collected before (round 0), during (round 6), and after (round 12) the recirculation experiment, both at rest and after additional activation with PMA (panels a & b) or TRAP (panels c & d). BT (red symbols) represent blood samples set aside for the duration of each recirculation experiment to assess the effect of time alone on cell activation. Statistical significance is denoted by * for p < 0.05. Each symbol represents a different recirculation experiment (n = 5, using blood from 4 unique subjects).

WBCs in recirculating blood become progressively more activated, but at the end of the recirculation experiment (after round 12), they were no more activated than the BT control (Fig. 5a). WBCs removed with the retentate were about as activated as WBCs remaining in the recirculating blood, and there was no significant difference in activation between cells extracted at rounds 6 and 12 (Fig. 5b). Both recirculating and removed WBCs were able to become significantly more activated after incubation with PMA (positive control) regardless of when they were sampled during the recirculation experiment, which suggests that the cells did not become refractory after processing (Fig. 5a, b).

Despite the potential effect of shear on PLTs, activation of PLTs in the recirculating blood did not increase significantly over the entire duration of the experiment. After round 12 of recirculation through the CIF device, the level of PLT activation was similar to that of the BT control (Fig. 5c). Low activation could be caused by PLT refractoriness, which was not the case because recirculating PLTs were able to activate significantly upon exposure to TRAP (positive control) (Fig. 5c). Given the high sensitivity of PLTs to shear and our previously published findings with similar CIF devices43,44,45 we expected to observe at least some increase in activation of recirculating PLTs. One possible explanation is that most of the PLTs that became activated got promptly bound to the available WBCs to form PLT-WBC aggregates. Indeed, the number of PLT-WBC aggregates increased steadily and significantly over time during each recirculation experiment (Fig. 5d). PLTs participating in the PLT-WBC aggregates are excluded when PLT activation and count are measured, which can explain the activation of recirculating PLTs remaining relatively low (Fig. 5c) and their concentration gradually declining over time (Fig. 4b). Overall, the number of PLT-WBC aggregates in the recirculating blood after 12 rounds of recirculation was the same as in the BT control (Fig. 5d), suggesting that the increase was likely due to processing duration, and contribution from the CIF device was relatively minor. It further follows that one potential solution for minimizing the cell activation and PLT-WBC aggregate formation would be to reduce processing duration by increasing either the volumetric throughput (e.g., via additional multiplexing) or the separation efficiency (e.g., via additional design improvements) of the device.

Finally, we tested whether recirculation through the CIF device damaged RBCs by measuring free hemoglobin (Hb) and potassium levels before (round 0), during (round 6), and immediately after (round 12) each experiment. We previously employed the same assays to sensitively identify sublethal damage caused by centrifugation to stored RBCs during washing47,48. Throughout the CIF recirculation procedure, we did not observe any significant changes in free Hb levels and potassium levels remained below the detectable level (< 0.2 mM) in all samples (Table S1). Absent any indication of hemolysis, the most likely explanation for the observed decline of RBC concentration in the recirculating blood (Fig. 4b) is that an excessive number of RBCs was removed with the retentate. Indeed, the RBC concentration in the retentate (waste) was consistently higher than in the recirculating blood (Fig. 4b), on average by about 0.1 × 106/µL (Table S2). As discussed in our earlier study in which we observed a similar buildup of RBCs in the retentate when processing leukapheresis samples46, RBCs are highly deformable biconcave discs (~ 8 µm in diameter and only ~ 2 µm thick) and therefore their effective diameter could vary widely depending on the specific conditions near each filtration gap. Most RBCs have an effective diameter sufficiently small to follow the filtrate, but some do not and therefore tend to remain in the central channel creating the characteristic buildup46.

Conventional apheresis machines utilize centrifugation to separate WBCs from RBCs and PLTs, which limits the degree to which the ECV of the leukapheresis circuit could be reduced to match the needs of pediatric patients. The void volume of the CIF device (0.4 mL) described in this study was at least 100-fold smaller than that of a typical centrifugation-based apheresis machine (150–250 mL)49. This dramatic reduction of the ECV represents the main advantage of using microfluidic cell separation to ultimately enable safe and effective leukapheresis procedure in young children. This study builds upon our earlier CIF designs46, to tackle an entirely novel, much more challenging application than what we have explored previously. For the first time, a CIF microfluidic device is demonstrated capable of separating WBCs from diluted whole blood with > 80% efficiency while operating in the recirculation regime for > 3 hours without a noticeable decline in separation performance and with minimal effect on cell properties. Additionally, the current device operated optimally at an HCT that was 2 times higher, and with RBC and PLT loss that was 2–3 times lower, than in any of our previous reports46.

Pediatric patients undergoing leukapheresis for leukodepletion or cellular collection are typically anemic, with an average HCT of 20–30%49,50. At 20% HCT, the CIF device had a WBC removal efficiency of ~ 60%, which may be an acceptable level of performance in some patients. To operate the current CIF device prototype at its peak separation efficiency (WBC removal > 80%), blood would have to be diluted to 10% HCT before entering the device and then concentrated back to its native HCT before returning to the patient. Such hemoconcentration is performed routinely during pediatric cardiopulmonary bypass and extracorporeal membrane oxygenation to remove excess fluids, using devices with excellent biocompatibility and minimal void volume (e.g., 8 mL, Hemocor HPH Junior, Minntech Corp., Minneapolis, MN) at flow rates compatible with leukapheresis (as high as 100 mL/min)51. The decline in separation performance of the device at higher HCTs could be explained by the increasing number of stochastic cell–cell interactions and the deviation of the apparent viscosity of blood in the channels from the model we used when designing the device, both of which become more pronounced with increasing HCT46,52,53. Additional research to address these factors will be needed to further increase the CIF separation efficiency at higher HCTs.

The flow rate at which the CIF device operated with peak separation performance in our study (10 mL/min) is similar to clinical leukapheresis procedures, which are typically performed at 10–50 mL/min1,17,54. The observed decline in the separation efficiency at higher flow rates was likely due to the deformation (bulging) of the device channels at higher driving pressures, as we previously observed for similar devices46. This effect could be minimized by ultimately fabricating CIF devices from hard thermoplastics rather than the PDMS elastomer we used in this study39,40,55.

The HCT (10%) and flow rate (10 mL/min) at which the CIF device operated with its optimal performance were significantly higher than previously reported for other microfluidic devices designed to separate WBCs from minimally diluted whole blood39,40,42,56,57,58,59,61. Importantly, this study was the first to demonstrate a microfluidic device capable of highly efficient separation of cells from blood while operating in the recirculation regime, as opposed to the flow-through regime employed by all other microfluidic cell separation devices32.

The recirculation in this study was accomplished using a syringe pump programmed to go through multiple ‘withdraw-infuse’ cycles over the entire duration of the procedure. Such an approach to recirculation closely resembles the operation of an apheresis machine in the discontinuous mode typically employed for larger ECV or longer duration procedures1,49, or the manual exchange transfusion sometimes performed for young children when centrifugation-based leukapheresis is considered too risky due to a patient's hemodynamic status, relatively small TBV compared to the ECV of the circuit, and/or ability to place an appropriately sized catheter28,62. Remarkably, the CIF device was able to process blood without any signs of clogging while maintaining its separation efficiency for > 3 h, which is the typical duration of a conventional leukapheresis procedure1,17,54,62. This finding is particularly significant because any device intended for leukapheresis must be able to process recirculating blood over an extended period of time.

In the context of leukoreduction, conventional centrifugation-based leukapheresis would typically reduce WBC concentration by about one-third when processing 1–2 TBV50 and by about one-half when processing 2–3 TBV63. The CIF device was able to reduce the WBC concentration in the recirculating blood about twice as fast—by one-half after processing just one TBV, and by four-fifths after processing two TBV. From the clinical perspective, minimizing the loss of RBCs and PLTs during the leukapheresis procedure is important for reducing the risks associated with transfusion of allogeneic blood products. The loss of PLTs during the CIF device operation was ~ 2–3 times lower than would be typical for a centrifugation-based procedure (for which PLTs are removed at the same rate as WBCs because both cell types co-localize in the ‘buffy coat’ layer)50,63. Additional research will be needed to increase the flow ratio of the CIF device to further reduce the RBC and PLT losses.

Exposure to high shear forces during centrifugation can cause mechanical damage to cell membranes, induce excessive cell activation, and trigger hemostatic responses, contributing to the host of adverse outcomes associated with leukapheresis in neonates and infants21,24,28,64. In our experiments, WBCs and PLTs in the recirculating blood were no more activated than the cells in the BT control (sample left on the bench for the duration of the procedure), suggesting minimal additional activation contributed by the CIF device. Additionally, both WBCs and PLTs were able to become activated after exposure to an appropriate stimulant, suggesting that the cells were not refractory after CIF processing. Finally, we found no evidence of damage to the RBCs either. Taken together, our data strongly suggest that the CIF device operating in the closed-loop recirculation regime did not significantly activate or damage the blood cells.

Centrifugation is used to perform leukapheresis because it has been the only technology capable of separating WBCs with the required efficiency and volumetric throughput from recirculating blood. This study demonstrates the feasibility of using high-throughput microfluidic cell separation technology to ultimately enable centrifugation-free, low-ECV leukapheresis. Such a capability would be particularly useful in young children, a vulnerable group of patients who are currently underserved.

All experiments were performed in accordance with guidelines and regulations established by the University of Houston and the U.S. Department of Health and Human Services for the protection of human subjects. All experimental protocols involving human blood samples were approved by the University of Houston Institutional Review Board (Committee for the Protection of Human Subjects 1, protocol #16272-01). Informed consent was obtained from all subjects and/or their legal guardian(s). Units of whole blood (WB) were purchased from the Gulf Coast Regional Blood Center (Houston, TX). Samples of fresh WB were obtained by venipuncture from healthy volunteers (anticoagulant: acid citric dextrose, solution A; Vacutainer, BD Biosciences, Franklin Lakes, NJ). Samples were used immediately or stored in a blood bank refrigerator (4 °C, iB111, Helmer Scientific, Noblesville, IN) until use, and diluted with isotonic saline (0.9% w/v NaCl, RICCA Chemical Company, Arlington, TX) to achieve the desired hematocrit (HCT).

The design and fabrication of devices based on controlled incremental filtration (CIF) technology have been previously described in detail43,44,45,46. In a typical CIF design, the width of all filtration gaps is fixed at ~ 20 µm because microfluidic devices with a ‘minimum feature size’ less than 20 µm are very difficult to mass produce using currently available manufacturing methods43. Similarly, manufacturability of microfluidic devices with aspect ratio (depth:width) above ~ 7:1 is very limited. Therefore, CIF devices are designed to have channel depths of ~ 140 µm to have fluidic resistance as low as possible (and therefore higher flow rate/throughput at a given driving pressure) while still maintaining the fundamental manufacturability of the design. Given that the width of the gaps is fixed, the fraction of flow passing from the central to the side channels at each gap is controlled by the change in the widths of the channels, as described previously in detail44. If this filtration flow fraction is zero, no fluid is flowing through the gaps, and therefore no cells are carried by the fluid into the side channels. As the filtration flow fraction increases, the amount of fluid flowing through each gap increases, and therefore the size of the cells that are sufficiently small to be pulled by the flow into the side channels also increases. Cells that are too large to be pulled by the fluid flowing through the gaps will remain in the central channel. Thus, the magnitude of the fluid flow through the gaps determines the separation size cutoff (not the gap width which remains constant). This decoupling of the size cutoff from the gap width allows CIF devices with 20 µm gaps to successfully separate particles/cells that are smaller than the gaps, as demonstrated previously43,44,45,46.

The CIF device design, generated using custom code in MATLAB (The MathWorks Inc, Natick, MA), was transferred into a ~ 140 µm-thick layer of photoresist (SU-8 3050; Kayaku Advanced Materials Inc, Westborough, MA) on a 4″ silicon wafer (University Wafer, South Boston, MA) using soft lithography. The master wafer was replicated in poly(dimethylsiloxane) (PDMS; Sylgard 184, Dow Corning Corp, Midland, MI), and the PDMS replica (device layer) was sealed against a PDMS-coated Petri dish (flat substrate) using oxygen plasma (PDC-001, Harrick Plasma, Ithaca, NY). The inlet and outlet ports in the device layer were created using biopsy punches (Acuderm Inc, Fort Lauderdale, FL) of the appropriate size to match tubing connections (1.02- and 0.58-mm inner diameters; Scientific Commodities, Havasu City, AZ). An additional PDMS layer containing a system of large channels for collecting the filtrate and retentate outputs from the individual CIF elements of a multiplexed device was bonded on top of the device layer. After bonding, each assembled CIF device was treated with 1% (w/v) aqueous solution of mPEG-silane (MW 5000, Laysan Bio Inc, Arab, AL) for 25 min at 70 °C. Finally, the device was flushed with GASP buffer (9 mM Na2HPO4, 1.3 mM NaH2PO4, 140 mM NaCl, 5.5 mM glucose, 1% w/v bovine serum albumin, 290 mmol/kg, pH 7.4) and stored at 4 °C until use.

The recirculation setup consisted of a blood bag (500 mL; Fenwal 4R1590, GenesisBPS, Ramsey, NJ) which was connected to the multiplexed CIF device and other components of the circuit by plastic tubing (Scientific Commodities) linked through appropriately sized Luer-lock connectors (Qosina, Ronkonkoma, NY). The tubing conveying the blood sample from the bottom of the blood bag to the CIF device was connected to a syringe pump (Genie Touch, Kent Scientific, Torrington, CT) and the inlet of the CIF device via two 3-way stopcocks (Qosina). One stopcock was used for sampling the blood coming from the bag, and the other stopcock was used for connecting the syringe pump with either the outlet of the bag (to withdraw blood from the bag) or the inlet of the device (to infuse blood through the device). The mode of pump operation (infuse/withdraw) and the position of the stopcock were set manually during the experiment. The blood bag was mixed by hand during and after each recirculation round.

Complete blood counts with a 5-part differential were measured using a hematology analyzer (XS-1000i, Sysmex America, Inc., Mundelein, IL). WBC, RBC and PLT counts were used to calculate the percent removal for each cell type as follows: \(\% removal={C}_{r}/\left({C}_{r}+{C}_{f}\times FR\right)\times 100\), where \({C}_{r}\) is the cell count in the retentate (central channel output of all CIF elements), \({C}_{f}\) is the cell count in the filtrate (side channel output of all CIF elements), and \(FR\) is the flow ratio of the device (defined as the ratio of the cumulative volume of the filtrate output to that of the retentate output of the device)46. All flow ratios were calculated based on measured volumes.

Imaging flow cytometry (FC, Amnis ImagestreamX Mk II, Luminex Corporation, Austin, TX) was used to measure PLT and WBC surface antigen activation markers, and the prevalence of PLT-WBC aggregates, using the following antibody cocktails (all from BD Biosciences, San Jose, CA). PLT activation: CD41a/APC (20 µL; BD 559,777), CD62P/PE (20 µL; BD 555,524), Dulbecco's phosphate-buffered saline without calcium or magnesium (DPBS-/-; 10 µL). Thrombin receptor agonist peptide 6 (TRAP; 70 µM; Sigma) was used as the positive control for PLT activation. WBC activation: CD45/APC (5 µL; BD 561,864), CD62L/FITC (20 µL; BD 555,543), CD11b/PE (20 µL; BD 555,388), and DPBS-/- (25 µL). Phorbol myristate acetate (PMA; 0.5 µg/mL; Sigma) was used as the positive control for WBC activation. PLT-WBC aggregates (PLA): CD45/APC (5 µL), CD41a/FITC (20 µL; BD 555,466), and DPBS-/- (45 µL). TRAP (70 µM) was used as the positive control for PLA formation. To perform the FC measurements, 30 µL of the blood sample were added to each antibody cocktail, gently mixed, and left to incubate in the dark for 15 min at room temperature (RT). After antibody labeling, RBCs were lysed using 1X BD FACS lysing solution for 15 min at RT. After lysis, samples were centrifuged at 900 × g for 5 min to remove the supernatant, pelleted cells were resuspended in 100 µL of 1% paraformaldehyde and then stored at 4ºC until FC measurements were performed (within 24 h).

The level of free hemoglobin (Hb) in the supernatant was measured using the modified cyanmethemoglobin method following the manufacturer's instructions (Drabkin's reagent; D5941, Sigma). Briefly, a blood sample was centrifuged at 1000 × g for 5 min to pellet the RBCs, then 40 µL of the supernatant was added to 160 µL of Drabkin's reagent, and incubated for 20 min. The absorbance was measured at 540 nm using a plate reader (SpectraMax M5, Molecular Devices, Sunnyvale, CA). The concentration of Hb was calculated using a calibration curve constructed with a human Hb standard (Pointe Scientific Inc, Canton MI). The level of potassium (K+) was measured using a handheld blood analyzer (i-STAT, Abbott Laboratories, Abbott Park, IL) using CHEM8 + cartridges, as previously described47.

All values were expressed as mean ± standard deviation. Statistical significance (defined as p < 0.05) of the observed differences was determined using the paired two-sided t-test for cell count data, one-way ANOVA for Hb measurements, and either 2-way repeated measures ANOVA or a mixed-effects model (restricted maximum likelihood) matched for both time and activation state with Sidak's multiple comparisons test for FC measurements.

Ganzel, C., Becker, J., Mintz, P. D., Lazarus, H. M. & Rowe, J. M. Hyperleukocytosis, leukostasis and leukapheresis: Practice management. Blood Rev. 26, 117–122. https://doi.org/10.1016/j.blre.2012.01.003 (2012).

Article PubMed Google Scholar

Padmanabhan, A. Cellular collection by apheresis. Transfusion 58(Suppl 1), 598–604. https://doi.org/10.1111/trf.14502 (2018).

Article PubMed Google Scholar

Blum, W. & Porcu, P. Therapeutic apheresis in hyperleukocytosis and hyperviscosity syndrome. Semin. Thromb. Hemost. 33, 350–354. https://doi.org/10.1055/s-2007-976170 (2007).

Article PubMed Google Scholar

Röllig, C. & Ehninger, G. How I treat hyperleukocytosis in acute myeloid leukemia. Blood 125, 3246–3252 (2015).

Article Google Scholar

Sigurbjornsson, F. T. & Bjarnason, I. Leukocytapheresis for the treatment of IBD. Nat. Clin. Pract. Gastroenterol. Hepatol. 5, 509–516. https://doi.org/10.1038/ncpgasthep1209 (2008).

Article PubMed Google Scholar

Hanai, H. et al. The mode of actions of the Adacolumn therapeutic leucocytapheresis in patients with inflammatory bowel disease: A concise review. Clin. Exp. Immunol. 163, 50–58. https://doi.org/10.1111/j.1365-2249.2010.04279.x (2011).

Article CAS PubMed PubMed Central Google Scholar

Saniabadi, A. R., Tanaka, T., Yamamoto, T., Kruis, W. & Sacco, R. Granulomonocytapheresis as a cell-dependent treatment option for patients with inflammatory bowel disease: Concepts and clinical features for better therapeutic outcomes. J. Clin. Apher. 34, 51–60. https://doi.org/10.1002/jca.21670 (2019).

Article PubMed Google Scholar

Takakura, M. et al. Successful treatment of rituximab- and steroid-resistant nephrotic syndrome with leukocytapheresis. J. Clin. Apher. 33, 409–411. https://doi.org/10.1002/jca.21613 (2018).

Article PubMed Google Scholar

Miyazono, A. et al. Successful remission induced by plasma exchange combined with leukocytapheresis against refractory systemic juvenile idiopathic arthritis. Eur. J. Pediatr. 173, 1557–1560. https://doi.org/10.1007/s00431-013-2093-5 (2014).

Article PubMed Google Scholar

Nikolajeva, O. et al. Single-donor granulocyte transfusions for improving the outcome of high-risk pediatric patients with known bacterial and fungal infections undergoing stem cell transplantation: a 10-year single-center experience. Bone Marrow Transplant. 50, 846–849. https://doi.org/10.1038/bmt.2015.53 (2015).

Article CAS PubMed Google Scholar

Price, T. H. et al. Efficacy of transfusion with granulocytes from G-CSF/dexamethasone-treated donors in neutropenic patients with infection. Blood 126, 2153–2161. https://doi.org/10.1182/blood-2015-05-645986 (2015).

Article CAS PubMed PubMed Central Google Scholar

Nair, S. K. et al. Ex vivo generation of dendritic cells from cryopreserved, post-induction chemotherapy, mobilized leukapheresis from pediatric patients with medulloblastoma. J. Neurooncol. 125, 65–74. https://doi.org/10.1007/s11060-015-1890-2 (2015).

Article CAS PubMed PubMed Central Google Scholar

Maude, S. L. et al. Tisagenlecleucel in children and young adults with B-cell lymphoblastic leukemia. N. Engl. J. Med. 378, 439–448. https://doi.org/10.1056/NEJMoa1709866 (2018).

Article CAS PubMed PubMed Central Google Scholar

Ceppi, F. et al. Lymphocyte apheresis for chimeric antigen receptor T-cell manufacturing in children and young adults with leukemia and neuroblastoma. Transfusion 58, 1414–1420. https://doi.org/10.1111/trf.14569 (2018).

Article CAS PubMed Google Scholar

Cerwenka, A. & Lanier, L. L. Natural killers join the fight against cancer. Science 359, 1460–1461. https://doi.org/10.1126/science.aat2184 (2018).

Article ADS CAS PubMed Google Scholar

Diaz, M. A. et al. Large-volume leukapheresis in pediatric patients: pre-apheresis peripheral blood CD34+ cell count predicts progenitor cell yield. Haematologica 84, 32–35 (1999).

CAS PubMed Google Scholar

Cecyn, K. Z. et al. Large-volume leukapheresis for peripheral blood progenitor cell collection in low body weight pediatric patients: a single center experience. Transfus. Apher. Sci. 32, 269–274. https://doi.org/10.1016/j.transci.2004.07.015 (2005).

Article PubMed Google Scholar

Karow, A. et al. Peripheral blood progenitor cell collection in pediatric patients optimized by high pre-apheresis count of circulating CD34+ cells and high blood flow. Bone Marrow Transplant. 54, 885–893. https://doi.org/10.1038/s41409-018-0353-8 (2019).

Article CAS PubMed Google Scholar

Ribeil, J. A. et al. Gene therapy in a patient with sickle cell disease. N. Engl. J. Med. 376, 848–855. https://doi.org/10.1056/NEJMoa1609677 (2017).

Article CAS PubMed Google Scholar

Thompson, A. A. et al. Gene therapy in patients with transfusion-dependent beta-thalassemia. N. Engl. J. Med. 378, 1479–1493. https://doi.org/10.1056/NEJMoa1705342 (2018).

Article CAS PubMed Google Scholar

Pathiraja, V., Matar, A. J., Gusha, A., Huang, C. A. & Duran-Struuck, R. Leukapheresis protocol for nonhuman primates weighing less than 10 Kg. J. Am. Assoc. Lab. Anim. Sci. 52, 70–77 (2013).

CAS PubMed PubMed Central Google Scholar

Rudolph, C. D. Rudolph's pediatrics (Appleton & Lange, 2003).

Google Scholar

Sisson, T. R., Whalen, L. E. & Telek, A. The blood volume of infants. II. The premature infant during the first year of life. J. Pediatr. 55, 430–446. https://doi.org/10.1016/s0022-3476(59)80281-x (1959).

Article CAS PubMed Google Scholar

Maitta, R. W. Current state of apheresis technology and its applications. Transfus. Apher. Sci. 57, 606–613. https://doi.org/10.1016/j.transci.2018.09.009 (2018).

Article PubMed Google Scholar

Michon, B. et al. Complications of apheresis in children. Transfusion 47, 1837–1842. https://doi.org/10.1111/j.1537-2995.2007.01405.x (2007).

Article PubMed Google Scholar

McLeod, B. C. et al. Frequency of immediate adverse effects associated with therapeutic apheresis. Transfusion 39, 282–288. https://doi.org/10.1046/j.1537-2995.1999.39399219285.x (1999).

Article CAS PubMed Google Scholar

Woloskie, S., Armelagos, H., Meade, J. M. & Haas, D. Leukodepletion for acute lymphocytic leukemia in a three-week-old infant. J. Clin. Apher. 16, 31–32. https://doi.org/10.1002/jca.1006 (2001).

Article CAS PubMed Google Scholar

Runco, D. V. et al. Hyperleukocytosis in infant acute leukemia: A role for manual exchange transfusion for leukoreduction. Transfusion 58, 1149–1156. https://doi.org/10.1111/trf.14512 (2018).

Article PubMed PubMed Central Google Scholar

Neron, S., Dussault, N. & Racine, C. Whole-blood leukoreduction filters are a source for cryopreserved cells for phenotypic and functional investigations on peripheral blood lymphocytes. Transfusion 46, 537–544. https://doi.org/10.1111/j.1537-2995.2006.00772.x (2006).

Article PubMed Google Scholar

de Amorim, C. G. et al. Leukocyte depletion during CPB: Effects on inflammation and lung function. Inflammation 37, 196–204. https://doi.org/10.1007/s10753-013-9730-z (2014).

Article CAS PubMed Google Scholar

Sethu, P., Sin, A. & Toner, M. Microfluidic diffusive filter for apheresis (leukapheresis). Lab Chip 6, 83–89. https://doi.org/10.1039/b512049g (2006).

Article CAS PubMed Google Scholar

Warkiani, M. E., Wu, L., Tay, A. K. & Han, J. Large-volume microfluidic cell sorting for biomedical applications. Annu. Rev. Biomed. Eng. 17, 1–34. https://doi.org/10.1146/annurev-bioeng-071114-040818 (2015).

Article CAS PubMed Google Scholar

Martel, J. M. & Toner, M. Inertial focusing in microfluidics. Annu. Rev. Biomed. Eng. 16, 371–396. https://doi.org/10.1146/annurev-bioeng-121813-120704 (2014).

Article CAS PubMed PubMed Central Google Scholar

Di Carlo, D. Inertial microfluidics. Lab Chip 9, 3038–3046. https://doi.org/10.1039/b912547g (2009).

Article CAS PubMed Google Scholar

Nivedita, N. & Papautsky, I. Continuous separation of blood cells in spiral microfluidic devices. Biomicrofluidics 7, 54101. https://doi.org/10.1063/1.4819275 (2013).

Article CAS PubMed Google Scholar

Zhang, J. et al. High-throughput separation of white blood cells from whole blood using inertial microfluidics. IEEE Trans. Biomed. Circuits Syst. 11, 1422–1430. https://doi.org/10.1109/TBCAS.2017.2735440 (2017).

Article PubMed Google Scholar

Jeon, H. et al. Fully-automated and field-deployable blood leukocyte separation platform using multi-dimensional double spiral (MDDS) inertial microfluidics. Lab Chip 20, 3612–3624. https://doi.org/10.1039/d0lc00675k (2020).

Article CAS PubMed Google Scholar

McGrath, J., Jimenez, M. & Bridle, H. Deterministic lateral displacement for particle separation: A review. Lab Chip 14, 4139–4158. https://doi.org/10.1039/c4lc00939h (2014).

Article CAS PubMed Google Scholar

Martel, J. M. et al. Continuous flow microfluidic bioparticle concentrator. Sci. Rep. 5, 11300. https://doi.org/10.1038/srep11300 (2015).

Article ADS CAS PubMed PubMed Central Google Scholar

Mutlu, B. R. et al. Non-equilibrium inertial separation array for high-throughput, large-volume blood fractionation. Sci. Rep. 7, 9915. https://doi.org/10.1038/s41598-017-10295-0 (2017).

Article ADS CAS PubMed PubMed Central Google Scholar

D’Silva, J., Austin, R. H. & Sturm, J. C. Inhibition of clot formation in deterministic lateral displacement arrays for processing large volumes of blood for rare cell capture. Lab Chip 15, 2240–2247. https://doi.org/10.1039/c4lc01409j (2015).

Article CAS PubMed PubMed Central Google Scholar

Campos-Gonzalez, R. et al. Deterministic lateral displacement: the next-generation CAR T-cell processing?. SLAS Technol. 23, 338–351. https://doi.org/10.1177/2472630317751214 (2018).

Article CAS PubMed Google Scholar

Gifford, S. C., Spillane, A. M., Vignes, S. M. & Shevkoplyas, S. S. Controlled incremental filtration: a simplified approach to design and fabrication of high-throughput microfluidic devices for selective enrichment of particles. Lab Chip 14, 4496–4505. https://doi.org/10.1039/c4lc00785a (2014).

Article CAS PubMed PubMed Central Google Scholar

Xia, H., Strachan, B. C., Gifford, S. C. & Shevkoplyas, S. S. A high-throughput microfluidic approach for 1000-fold leukocyte reduction of platelet-rich plasma. Sci. Rep. 6, 35943. https://doi.org/10.1038/srep35943 (2016).

Article ADS CAS PubMed PubMed Central Google Scholar

Gifford, S. C. et al. A portable system for processing donated whole blood into high quality components without centrifugation. PLoS One 13, e0190827. https://doi.org/10.1371/journal.pone.0190827 (2018).

Article CAS PubMed PubMed Central Google Scholar

Strachan, B. C., Xia, H., Voros, E., Gifford, S. C. & Shevkoplyas, S. S. Improved expansion of T cells in culture when isolated with an equipment-free, high-throughput, flow-through microfluidic module versus traditional density gradient centrifugation. Cytotherapy 21, 234–245. https://doi.org/10.1016/j.jcyt.2018.12.004 (2019).

Article CAS PubMed PubMed Central Google Scholar

Voros, E., Lu, M., Lezzar, D. & Shevkoplyas, S. S. Centrifugation-free washing reduces buildup of potassium and free hemoglobin in washed red blood cells after the procedure. Am. J. Hematol. 93, E389–E391. https://doi.org/10.1002/ajh.25277 (2018).

Article PubMed PubMed Central Google Scholar

Voros, E., Piety, N. Z., Strachan, B. C., Lu, M. & Shevkoplyas, S. S. Centrifugation-free washing: A novel approach for removing immunoglobulin A from stored red blood cells. Am. J. Hematol. 93, 518–526. https://doi.org/10.1002/ajh.25026 (2018).

Article CAS PubMed PubMed Central Google Scholar

Jarisch, A. et al. Unstimulated apheresis for chimeric antigen receptor manufacturing in pediatric/adolescent acute lymphoblastic leukemia patients. J. Clin. Apher. 35, 398–405. https://doi.org/10.1002/jca.21812 (2020).

Article PubMed Google Scholar

Zeng, F. et al. Leukapheresis in 15 patients weighing 20kg or less: A single centre experience. Transfus. Apher. Sci. 56, 889–893. https://doi.org/10.1016/j.transci.2017.11.001 (2017).

Article PubMed Google Scholar

Wang, S., Palanzo, D. & Undar, A. Current ultrafiltration techniques before, during and after pediatric cardiopulmonary bypass procedures. Perfusion 27, 438–446. https://doi.org/10.1177/0267659112450061 (2012).

Article CAS PubMed Google Scholar

Fung, Y. C. Stochastic flow in capillary blood vessels. Microvasc. Res. 5, 34–48. https://doi.org/10.1016/s0026-2862(73)80005-6 (1973).

Article CAS PubMed Google Scholar

Inglis, D. W. & Herman, N. A scalable approach for high throughput branch flow filtration. Lab Chip 13, 1724–1731. https://doi.org/10.1039/c3lc50192b (2013).

Article CAS PubMed Google Scholar

Doberschuetz, N. et al. Mobilized peripheral blood stem cell apheresis via Hickman catheter in pediatric patients. Transfusion 59, 1061–1068. https://doi.org/10.1111/trf.15113 (2019).

Article CAS PubMed Google Scholar

Sollier, E., Murray, C., Maoddi, P. & Di Carlo, D. Rapid prototyping polymers for microfluidic devices and high pressure injections. Lab Chip 11, 3752–3765. https://doi.org/10.1039/c1lc20514e (2011).

Article CAS PubMed Google Scholar

Lissandrello, C., Dubay, R., Kotz, K. T. & Fiering, J. Purification of lymphocytes by acoustic separation in plastic microchannels. SLAS Technol. 23, 352–363. https://doi.org/10.1177/2472630317749944 (2018).

Article CAS PubMed Google Scholar

Cheng, Y., Ye, X., Ma, Z., Xie, S. & Wang, W. High-throughput and clogging-free microfluidic filtration platform for on-chip cell separation from undiluted whole blood. Biomicrofluidics 10, 014118. https://doi.org/10.1063/1.4941985 (2016).

Article PubMed PubMed Central Google Scholar

Li, X., Chen, W., Liu, G., Lu, W. & Fu, J. Continuous-flow microfluidic blood cell sorting for unprocessed whole blood using surface-micromachined microfiltration membranes. Lab Chip 14, 2565–2575 (2014).

Article CAS Google Scholar

Tan, J. K. S., Park, S. Y., Leo, H. L. & Kim, S. Continuous separation of white blood cells from whole blood using viscoelastic effects. IEEE Trans. Biomed. Circuits Syst. 11, 1431–1437. https://doi.org/10.1109/Tbcas.2017.2748232 (2017).

Article PubMed Google Scholar

Wu, Z., Chen, Y., Wang, M. & Chung, A. J. Continuous inertial microparticle and blood cell separation in straight channels with local microstructures. Lab Chip 16, 532–542 (2016).

Article CAS Google Scholar

Kim, B., Choi, Y. J., Seo, H., Shin, E. C. & Choi, S. Deterministic migration-based separation of white blood cells. Small 12, 5159–5168. https://doi.org/10.1002/smll.201601652 (2016).

Article CAS PubMed Google Scholar

Schwartz, J. et al. Guidelines on the use of therapeutic apheresis in clinical practice-evidence-based approach from the writing committee of the American Society for Apheresis: The seventh special issue. J. Clin. Apher. 31, 149–162. https://doi.org/10.1002/jca.21470 (2016).

Article PubMed Google Scholar

Thapa, N. et al. Therapeutic leukocytapheresis in infants and children with leukemia and hyperleukocytosis: A single institution experience. J. Clin. Apher. 33, 316–323. https://doi.org/10.1002/jca.21610 (2018).

Article PubMed Google Scholar

Lu, M., Lezzar, D. L., Voros, E. & Shevkoplyas, S. S. Traditional and emerging technologies for washing and volume reducing blood products. J. Blood Med. 10, 37–46. https://doi.org/10.2147/JBM.S166316 (2019).

Article CAS PubMed PubMed Central Google Scholar

Download references

Research reported in this publication was supported in part by the University of Houston High Priority Area Research Seed Grant (I0503554, PI: SSS) and by the National Heart, Lung, and Blood Institute of the National Institutes of Health under Award Number R01HL151858 (PI: SSS). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health, the U.S. Department of Veterans Affairs, or the United States Government.

Department of Biomedical Engineering, University of Houston, 3605 Cullen Blvd, Houston, TX, 77204-5060, USA

Dalia L. Lezzar, Ravin Huerta, Anton Mukhamedshin, Madeleine Lu & Sergey S. Shevkoplyas

Division of Pediatric Critical Care Medicine, Baylor College of Medicine, Houston, TX, 77030, USA

Fong W. Lam

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

D.L.L., F.W.L. and S.S.S. designed the study and analyzed the data. F.W.L. and S.S.S. obtained funding and supervised the project. D.L.L. and M.L. fabricated the microfluidic devices. D.L.L., F.W.L., R.H. and M.L. performed the experiments. A.M. assisted with data analysis and interpretation. D.L.L., F.W.L. and S.S.S. wrote the manuscript. All authors critically reviewed and approved the manuscript.

Correspondence to Sergey S. Shevkoplyas.

SSS is an inventor on U.S. Patent #9,789,235 "Separation and concentration of particles" describing the controlled incremental filtration technology, and part-owner of Halcyon Biomedical Incorporated, a company that would benefit from its commercialization. All other authors declare no conflict of interest.

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and Permissions

Lezzar, D.L., Lam, F.W., Huerta, R. et al. A high-throughput microfluidic device based on controlled incremental filtration to enable centrifugation-free, low extracorporeal volume leukapheresis. Sci Rep 12, 13798 (2022). https://doi.org/10.1038/s41598-022-16748-5

Download citation

Received: 02 March 2022

Accepted: 14 July 2022

Published: 13 August 2022

DOI: https://doi.org/10.1038/s41598-022-16748-5

Anyone you share the following link with will be able to read this content:

Sorry, a shareable link is not currently available for this article.

Provided by the Springer Nature SharedIt content-sharing initiative

By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.